Tools
RGA Abacus
Three ways into the abacus: by species, by mass number, or by fingerprint.
Reading a full spectrum? Spectrum Abacus →
Worked example: Water
Read from the reference base and preloaded. Base peak M18, four fragments as a percent of that peak, then the ratio test that settles the reading.
- M18H₂O⁺100%
- M17OH⁺23%
- M1H⁺3%
- M16O⁺/CH₄⁺1.1%
- M2H₂⁺1%
If M17/M18 ≈ 0.23, water is main contributor
- Ratio
- M17 / M18
- Expected
- 0.23 ± 0.05
- Reading
- If ~0.23, water is main source of M17
Source: NIST Chemistry WebBook, SRD 69 ; 70 eV electron impact ; normalised to base peak = 100. CAS 7732-18-5.
All sources for this reference base
Report a value
Select a mass
Sources · 7
- NIST Chemistry WebBook, Mass Spectra (70 eV electron impact).
- NIST Electron-Impact Cross Section Database.
- MKS Instruments, MicroVision 2 Residual Gas Analyzer : Operating Manual, 2019.
- Solvay, Fomblin PFPE product documentation (M69 / M119 / M169 triplet).
- DuPont / Chemours, Viton and Kalrez guides (thermal decomposition, M100 marker).
- Dow Corning / Shin-Etsu, PDMS data (M73 / M147 / M207 triplet, silicone migration).
- Gamma Vacuum / Agilent, Ion Pump manuals (noble-gas pumping, cathode sputtering).
Rated [A] primary standard data, [B] manufacturer documentation. 70 eV cracking patterns are nominal and vary 10-20% between instruments : validate against a known calibration gas before quantitative use.